Answer: The V(D)J primer sequences used for enrichment are specifically targeted towards the constant region of the T- or B-cell transcripts. This region is constant across multiple transcripts, which allows us to capture multiple variants with just a few specific primers. We use a nested-PCR enrichment method in order to increase specificity of the amplification product.
Please see the tables below for information on which gene each primer maps to:
Human TCR
Gene | Outer primer sequence | Inner primer sequence |
TRA | TGAAGGCGTTTGCACATGCA | AGTCTCTCAGCTGGTACACG |
TRB | TCAGGCAGTATCTGGAGTCATTGAG | TCTGATGGCTCAAACACAGC |
Human BCR
Gene | Outer primer sequence | Inner primer sequence |
IGHA |
CAGGGCACAGTCACATCCT |
GGGAAGTTTCTGGCGGTCA |
IGHD |
TGCTGGACCACGCATTTGTA | GGTGGTACCCAGTTATCAAGCAT |
IGHE |
GGTTTTGTTGTCGACCCAGTCT | GTGTCCCAGGTCACCATCAC |
IGHG |
TTGTCCACCTTGGTGTTGCT | TCCTGAGGACTGTAGGACAGC |
IGHM |
CATGACGTCCTTGGAAGGCA | CACGCTGCTCGTATCCGA |
IGLC |
TGTGGGACTTCCACTG | TAGCTGCTGGCCGC |
IGKC |
TTCTCGTAGTCTGCTTTGCTCAG | GCGTTATCCACCTTCCACTGT |
Mouse TCR
Gene | Outer primer sequence | Inner primer sequence |
TRA | CTGGTTGCTCCAGGCAATGG | AGTCAAAGTCGGTGAACAGGCA |
TRB | TGTAGGCCTGAGGGTCCGT | GGCCAAGCACACGAGGGTA |
Mouse BCR
Gene | Outer primer sequence | Inner primer sequence |
IGLC_Group1 | TCAGCACGGGACAAACTCTTCT | TACACACCAGTGTGGCCTT |
IGLC_Group2 | GCAGGAGACAGACTCTTCTCCA | CAGGCCACTGTCACACCACT |
IGHA_All | AACTGGCTGCTCATGGTGT | CAGGTCACATTCATCGTGCCG |
IGHD_All | TGGTGCAAGTGTGGTTGAGGT | GAGGCCAGCACAGTGACCT |
IGHE_Group1 | TGGTCACTTGGCTGGTGGTG | GCAGGGAAGTTCACAGTGCT |
IGHE_Group2 | CACTTGGCAGGTGAACTGTTTTCT | CTGTTTGAGATCAGTTTGCCATCCT |
IGHM_All | AACCTTCAAGGATGCTCTTGGGA | TGCGAGGTGGCTAGGTACTTG |
IGHG_Group1 | GGACAGGGATCCAGAGTTCCA | CCCTTGACCAGGCATCC |
IGHG_Group2 | AGGTGACGGTCTGACTTGGC | AGGTCACGGAGGAACCAGTTG |
IGHG_Group3 | GCTGGACAGGGCTCCATAGTT | GGCATCCCAGTGTCACCGA |
IGHG_Group4 | GGCACCTTGTCCAATCATGTTCC | AGAAGATCCACTTCACCTTGAAC |
IGKC_All | ATGTCGTTCATACTCGTCCTTGGT | GAAGCACACGACTGAGGCAC |