Question: If I have a wetting failure/clog in one of the sample wells of the HT assay, will this impact the other paired sample?
Answer: We have performed tests in-house to verify a failure (either wetting or clog) observed in the recovery well (north 3A) does not impact the paired recovery well (south 3B). In very rare instances, we have observed failures in both 3A and 3B. This only occurs in samples with very high debris when a complete clog occurs and triggers a failure in the paired well.
Products: Single Cell Gene Expression, Single Cell Immune Profiling