Question: What are the specifications of the Chromium X Series?
Answer: The specifications of the Chromium X Series are as follows:
Parameter |
Specification |
Weight |
41.4 lb (18.8 kg)* |
Instrument Dimensions with: Tray close Tray open |
W D H 12.2”[31.0 cm] × 18.6”[47.2 cm] × 10.2”[26.0 cm] 12.2”[31.0 cm] × 22.8”[58.0 cm] × 10.2”[26.0 cm] |
Electrical Requirements Nominal from a standard 3-prong wall receptacle that includes a safety ground pin |
100-240 VAC, 50/60 Hz, 250 W** 90-264 V operational range (+/- 10% of nominal) Overvoltage Category II (standard receptacle) |
Pollution |
Degree 2 (Indoor Use Only) |
Ventilation Requirement |
Minimum 4” [10 cm] Around all sides |
Operating Temperature |
64-82°F [18-28°C] Use in a typical indoor laboratory environment. Extreme temperature conditions will affect the sensitive reagents used with the instrument. |
Humidity |
30-80% R.H. non-condensing |
Altitude |
Altitude up to 7500 ft [2286 m] above sea level |
Power Cable Length Power cables will be in accordance with regional specifications |
6-7 ft [2 m] Standard |
* Shipping overpack box will be labeled “Team Lift” due to increased weight
**Typically consumes ~30 W, but can increase for a brief period under peak demand conditions (large temperature setpoint change)
Products: Chromium X Series