Question: Can I sequence Single Cell ATAC libraries with other libraries?
Answer: We do not recommend pooling Single Cell ATAC libraries with other non-ATAC 10x sequencing libraries. Sequencing ATAC libraries with other 10x single-indexed libraries (SC3', SC5') in the same or different lane is not supported as the dual indexing required for ATAC libraries may cause sequencing issues for these other libraries.
Pooling with 10x dual-indexed gene expression libraries (SC3', SC5') is also not recommended since the lengths of the i5 and i7 indices are different between ATAC and dual-indexed gene expression libraries.
Sequencing Single Cell ATAC libraries in the same lane as non-10x libraries are not supported, as we cannot guarantee that mixing libraries will not impact sequencing data quality for non-Single Cell ATAC libraries.
ATAC libraries can be sequenced with Multiome ATAC libraries, but only on specific sequencers. Please review this article carefully: Why can I not pool Multiome ATAC and Standalone ATAC libraries on certain sequencers?
Refer to the following articles for more details on altering read lengths.
- Can I change the sequencing read lengths of R1N and R2N?
- Can I change the sequencing read lengths of the i5 and i7 indices?
Products: Single Cell Gene Expression, Single Cell Immune Profiling, Single Cell ATAC, Single Cell CNV