Question: Can different gene expression panels be surveyed on the same Xenium Analyzer run?
Answer: Yes, two different panels can be surveyed on the same Xenium Analyzer run. The caveat is that both slides must be prepared with the same assay chemistry. For example, it is not possible to run the Xenium Analyzer with one slide processed with Xenium Slides & Sample Prep Reagents (PN-1000460) and a second slide processed with Xenium Prime Sample Preparation Reagents (PN-1000720). This is because Xenium v1 vs. Xenium Prime chemistry require a different number of instrument imaging cycles, plus there are assay-specific decoding reagents.
Information about run set-up and panel section can be found in the Xenium Analyzer User Guide.
Products: Xenium In Situ Gene Expression, Xenium Prime In Situ Gene Expression