Question: What are the specifications of the Chromium Controller and Single Cell Controller?
Answer: The specifications are listed below:
Weight: 12.5 lb (5.6 kg)
Size: (W X D X H) 7.9 x 10.3 x 6.4 in (20 x 26.3 x 16.4 cm)
Ventilation Requirement: Minimum 4 in (10 cm) around all sides
Vibration Requirement: Do not place on lab bench with equipment that vibrates (vortexer, centrifuge, etc.)
Electrical Requirements: Nominal 100-240 Vac, 50-60 Hz, 25 W, from a standard 3-prong wall receptacle that includes a safety ground pin
Energy consumption: 15 watts
Heat load: 52 BTU/hr
Usable Temperature Range: 18-28°C (64-82°F)
Humidity: 85% Max (Non-Condensing)
Altitude: 0 - 6560 ft (0 - 2000 m)
For additional information, see the Chromium Controller Specifications manual: https://support.10xgenomics.com/genome-exome/index/doc/specifications-chromium-controller-specifications
Products: Chromium Controller, Chromium Single Cell Controller